Korean Circ J. 2021;51(12):1001-1014.

A First-in-Man Clinical Evaluation of Sirolimus and Ascorbic Acid-Eluting Stent Systems: a Multicenter, Subject-Blinded, Randomized Study.

, , , , , ,

제1저자 임영효,윤지현 · 교신저자 김효수,박준규

소속 (6)
  1. Division of Cardiology, Department of Internal Medicine, Hanyang University College of Medicine, Seoul, Korea.
  2. CG Bio Co. Ltd., Seoul, Korea.
  3. Division of Cardiology, Cardiovascular Center, Korea University Anam Hospital, Seoul, Korea.
  4. Department of Cardiology, Yonsei University Wonju Severance Christian Hospital, Wonju, Korea.
  5. CG Bio Co. Ltd., Seoul, Korea. pjk23@cgbio.co.kr.
  6. Department of Internal Medicine and Cardiovascular Center, Seoul National University College of Medicine, Seoul, Korea. hyosoo@snu.ac.kr.

Abstract

BACKGROUND AND OBJECTIVES This clinical trial was conducted to evaluate the safety and efficacy of D+Storm™ drug-eluting stent (DES) and BioMatrix Flex™ DES.

METHODS This study was a multicenter, subject-single-blind, randomized, and confirmed comparative clinical trial. According to the inclusion criteria, those diagnosed with stable angina, unstable angina, silent ischemia, or non-ST-segment myocardial infarction were selected among patients with coronary artery stenosis as subjects. Among the subjects with 50% stenosis on coronary angiography, the experiment was performed on those who had a lesion with reference vessel 2.5-4.0 mm in diameter and ≤40 mm in length. The primary endpoint was an in-segment late loss and the secondary endpoints were in-stent late lumen loss, stent malapposition, the incidence of mortality, myocardial infarction, reoperation, and stent thrombosis at 36 weeks.

RESULTS 57 patients in the D+Storm™ DES group and 55 patients in the BioMatrix Flex™ DES group were enrolled in the study. Fifty-seven patients in the D+Storm™ DES group and Fifty-five patients in the BioMatrix Flex™ DES group were enrolled in the study. An average of in-segment late lumen loss was 0.08±0.13 mm in the D+Storm™ DES group and 0.14±0.32 mm in the BioMatrix Flex™ DES group with no significant difference between the 2 groups (p=0.879). In addition, there was no significant difference in adverse events between D+Storm™ DES and BioMatrix Flex™ DES.

CONCLUSIONS This study demonstrated the clinical effectiveness and safety of D+Storm™ DES implantation in patients with coronary artery disease over a 36-week follow-up period.

Copyright © 2021. The Korean Society of Cardiology.

Keywords

Absorbable implantsAscorbic acidCoronary artery diseaseDrug-eluting stentsSirolimus

Figures

Figure 1
Figure 1 The flow-chart of this study. DES = drug-eluting stent.

그림 출처: PubMed Central 공개 논문 (PMC8636762)

Conflict of interest statement

Jun-kyu Park and Ji Hyun Youn are researchers at CG Bio, a manufacturer of D+Storm DES. This clinical trial was funded by CG Bio. The other authors report no conflicts.

그 밖의 정보
출판 형태
Journal Article
저널
Korean circulation journal · ISSN 1738-5520
연구비
Medical Device Development Fund grant · Ministry of Health and Welfare HI16C0994