Feasibility and measurement stability of smartwatch-based cuffless blood pressure monitoring: A real-world prospective observational study.
제1저자 한민주, Young-Ro Lee, 박태영 · 교신저자 이해영
소속 (13)
- Department of Internal Medicine, Seoul National University College of Medicine, Seoul, South Korea.
- Department of Internal Medicine, Seoul National University Hospital, Seoul, South Korea.
- Department of Electrical and Computer Engineering, College of Engineering, Seoul National University, Seoul, South Korea.
- Department of Applied Statistics and Department of Statistics and Data Science, Yonsei University, Seoul, Korea.
- Division of Cardiology, Department of Internal Medicine, Bucheon St. Mary's Hospital, College of Medicine, The Catholic University of Korea, Seoul, South Korea.
- Division of Cardiology, Department of Internal Medicine, Ewha Woman's University Seoul Hospital, Seoul, South Korea.
- Medical Outpatient and Hypertension Clinic, ESH Hypertension Centre of Excellence, University Hospital Basel, Basel, Switzerland.
- Department of Internal Medicine, College of Medicine, Chungbuk National University, Cheongju, Korea.
- ESH Excellence Hypertension Center, Department of Internal Medicine, Geriatrics Section, Hospital Clinic, University of Barcelona, Barcelona, Spain.
- Division of Cardiology, Department of Medicine, University of Washington School of Medicine, Seattle, WA, USA.
- Hypertension Center STRIDE-7, National and Kapodistrian University of Athens, School of Medicine, Third Department of Medicine, Sotiria Hospital, Athens, Greece.
- Department of Internal Medicine, Seoul National University College of Medicine, Seoul, South Korea. hylee612@snu.ac.kr.
- Department of Internal Medicine, Seoul National University Hospital, Seoul, South Korea. hylee612@snu.ac.kr.
Abstract
Cuffless wearable devices are currently being developed for long-term monitoring of blood pressure (BP) in patients with hypertension and in apparently healthy people. This study evaluated the feasibility and measurement stability of smartwatch-based cuffless BP monitoring in real-world conditions. Users of the first smartwatch-based cuffless BP monitor approved in Korea (Samsung Galaxy Watch) were invited to upload their data from using the device for 4 weeks post calibration. A total of 760 participants (mean age 43.7 ± 11.9, 80.3% men) provided 35,797 BP readings (average monitoring 22 ± 4 days [SD]; average readings 47 ± 42 per participant [median 36]). Each participant obtained 1.5 ± 1.3 readings/day and 19.7% of the participants obtained measurements every day. BP showed considerable variability, mainly depending on the day and time of the measurement. There was a trend towards higher BP levels on Mondays than on other days of the week and on workdays than in weekends. BP readings taken between 00:00 and 04:00 tended to be the lowest, whereas those between 12:00 and 16:00 the highest. The average pre-post calibration error for systolic BP (difference in 7-day BP before and after calibration), was 6.8 ± 5.6 mmHg, and was increased with higher systolic BP levels before calibration. Smartwatch-based cuffless BP monitoring is feasible for out-of-office monitoring in the real-world setting. The stability of BP measurement post calibration and the standardization and optimal time interval for recalibration need further investigation.
© 2023. The Author(s), under exclusive licence to The Japanese Society of Hypertension.
Keywords
Blood Pressure.Cuffless Blood Pressure MeasurementSmartwatchWearable Device