J Korean Med Sci. 2010;25(10):1462-6.

Long-term prognosis of left ventricular lead.

, , , , , , , , ,

제1저자 박승정 · 교신저자 오세일

소속 (1)
  1. Department of Internal Medicine, Seoul National University College of Medicine, Seoul, Korea.

Abstract

Transvenous left ventricular (LV) lead implantation is on the increase due to cardiac resynchronization therapy (CRT). However, there has been paucity of data on the prognosis of LV lead. Consecutive 32 patients with LV lead for CRT (n=22) or pacemaker (n=10) were subjected. Serial changes in pacing threshold and impedance along with lead-related complications were evaluated. Over 2 yr follow-up, there was no significant change in relative threshold voltage to the initial value (100%, 110%, 89.6%, and 79.6% at baseline, 1, 6, and 24 months respectively, P=0.62) as well as lead impedance (816±272, 650±178, 647±191, and 590±185 ohm at baseline, 1, 6, and 24 months respectively, P=0.80). The threshold change was not affected by lead position, lead polarity, and indication of lead implantation. The cumulative rates of lead revision were 6.3% (n=2) and 9.4% (n=3) in 6 month and 2 yr follow-up, respectively. One case of phrenic nerve capture at left lateral decubitus position was detected 1 month after the implantation. However, there were no serious complications over 2 yr period. In conclusion, transvenous LV lead implantation showed favorable long-term prognosis. Pacing parameters remained stable without significant changes over 2 yr follow-up.

Keywords

Cardiac Resynchronization TherapyCardiac veinPacemakerTricuspid Valve Replacement

Figures

Fig. 1
Fig. 1 Long-term stability of a left ventricular lead implanted in lateral cardiac veins. LV lead implantation was performed due to both complete atrioventricular block and mechanical tricuspid valve (A, B). The lead remained stable without dislodgement for 42 months (C). mAV, mechanical aortic valve; mMV, mechanical mitral valve; mTV, mechanical tricuspid valve.
Fig. 2
Fig. 2 Pacing parameters remain stable without significant changes over 2 yr follow-up. There was no significant change in relative threshold voltage to the initial value as well as lead impedance over 2 yr follow-up. *Relative threshold (%)=follow-up voltage/initial voltage×100.
Fig. 3
Fig. 3 Serial changes in LV pacing threshold and impedance in various subgroups. Long-term stabilities in threshold voltage and lead impedance are still maintained in various subgroups divided depending on lead position, lead polarity, and indication of lead implantation.

그림 출처: PubMed Central 공개 논문 (PMC2946656)

그 밖의 정보
출판 형태
Evaluation Study, Journal Article
저널
Journal of Korean medical science · eISSN 1598-6357
전자 출판
2010-09-17
MeSH
Adult · Aged · Analysis of Variance · Cardiac Electrophysiology · Cardiac Resynchronization Therapy · Cardiac Resynchronization Therapy Devices* · Female · Follow-Up Studies · Heart Ventricles* · Humans · Male · Middle Aged · Multivariate Analysis · Prognosis · Ventricular Dysfunction, Left / therapy