Circ Cardiovasc Interv. 2023;16(1):e012307.

BioMatrix Versus Orsiro Stents for Coronary Artery Disease: A Multicenter, Randomized, Open-Label Study.

, , , , , , , , , , , , , , , ,

제1저자 권주성, 윤창환 · 교신저자 채인호

소속 (10)
  1. Cardiovascular Center, Department of Internal Medicine, Seoul National University Bundang Hospital, Republic of Korea (C.-H.Y., J.-S.K., J.J.P., S.-H.K., S.-H.Kim, J.-W.S., T.-J.Y., I.-H.C.).
  2. Sejong General Hospital, Bucheon, Republic of Korea (Y.J.C.).
  3. Catholic Kwandong University International St Mary's Hospital, Incheon, Republic of Korea (M.-K.K.).
  4. Pusan National University Hospital, Republic of Korea (K.S.C.).
  5. Wonju Severance Hospital, Yonsei University College of Medicine, Republic of Korea (S.-H.L.).
  6. Gangnam Severance Hospital, Yonsei University College of Medicine, Seoul, Republic of Korea (B.K.H.).
  7. Korea University Guro Hospital, Seoul, Republic of Korea (S.-W.R.).
  8. Gachon University Gil Medical Center, Incheon, Republic of Korea (W.C.K.).
  9. Chungnam National University Hospital, Daejeon, Republic of Korea (J.-H.L.).
  10. Seoul Metropolitan Government Seoul National University Boramae Medical Center, Seoul, Republic of Korea (S.-H.K.).

Abstract

BACKGROUND Comparative studies of ultrathin-strut biodegradable polymer sirolimus-eluting stent (BP-SES) have reported promising results and validated its excellent outcomes in terms of safety and efficacy. However, there are limited studies comparing BP drug-eluting stents with struts of different thicknesses. We compared the long-term clinical outcomes of patients treated with an ultrathin-strut BP-SES or a thick-strut biodegradable polymer biolimus-eluting stent (BP-BES).

METHODS The BIODEGRADE trial (Biomatrix and Orsiro Drug-Eluting Stents in Angiographic Result in Patients With Coronary Artery Disease) is a multicenter prospective randomized study comparing coronary revascularization in patients with ultrathin-strut BP-SES and thick-strut BP-BES with the primary end point of target lesion failure at 18 months posttreatment. We performed the prespecified analysis of 3-year clinical outcomes.

RESULTS In total, 2341 patients were randomized to receive treatment with ultrathin-strut BP-SES (N=1175) or thick-strut BP-BES (N=1166). The 3-year incidence rate of target lesion failure was 3.2% for BP-SES and 5.1% for BP-BES (P=0.023). The difference was primarily due to differences in ischemia-driven target lesion revascularization (BP-SES, 1.5%; BP-BES, 2.8%; P=0.035) between groups. A landmark analysis of the late follow-up period showed significant differences in target lesion failure, with outcomes being better in BP-SES. Cardiac death and target lesion revascularization were significantly lower in the BP-SES group.

CONCLUSIONS In a large, randomized trial, the long-term clinical outcome of target lesion failure at 3 years was significantly better among patients treated with the ultrathin-strut BP-SES. The results indicate the superiority of the ultrathin-strut BP-SES compared with the thick-strut BP-BES.

REGISTRATION URL: https://clinicaltrials.gov; Unique identifier: NCT02299011.

Keywords

coronary artery diseasedrug-eluting stentischemiapercutaneous coronary interventionsirolimus

그 밖의 정보
출판 형태
Randomized Controlled Trial, Multicenter Study, Journal Article, Research Support, Non-U.S. Gov't
저널
Circulation. Cardiovascular interventions · eISSN 1941-7632
전자 출판
2022-12-07
MeSH
Humans · Coronary Artery Disease* / diagnostic imaging / therapy · Everolimus / adverse effects · Drug-Eluting Stents* · Prospective Studies · Treatment Outcome · Sirolimus / adverse effects · Polymers · Percutaneous Coronary Intervention* / adverse effects · Absorbable Implants · Prosthesis Design · Alkanesulfonic Acids
Substances
Everolimus · Sirolimus · Polymers · BES · Alkanesulfonic Acids
데이터
ClinicalTrials.gov: NCT02299011