Front Oncol. 2023;13:1103217.

Development of novel biliary metal stent with coil-spring structure and its application in vivo swine biliary stricture model.

, , , , , , , , , , , ,

제1저자 조인래 · 교신저자 이상협

소속 (6)
  1. Department of Internal Medicine and Liver Research Institute, Seoul National University College of Medicine, Seoul, Republic of Korea.
  2. Department of Gastroenterology, Kyung Hee University Hospital at Gangdong, Seoul, Republic of Korea.
  3. Department of Internal Medicine, Dongguk University College of Medicine, Dongguk University Ilsan Hospital, Goyang-si, Republic of Korea.
  4. Digestive Disease Center, Department of Internal Medicine, Inha University School of Medicine, Incheon, Republic of Korea.
  5. The National Center of Efficacy Evaluation for the Development of Health Products Targeting Digestive Disorders (NCEED) and Utah-Inha DDS & Advanced Therapeutics Research Center, Incheon, Republic of Korea.
  6. CG Bio Co., Ltd., Seoul, Republic of Korea.

Abstract

BACKGROUND As of date, endoscopic biliary stenting with plastic stent (PS) and self-expandable metal stent (SEMS) have been widely used for the palliation of biliary tract strictures. However, these two stents have several limitations regarding the management of biliary strictures caused by intrahepatic and hilar cholangiocarcinoma. PS has short patency and also risks bile duct injury and bowel perforation. SEMS is difficult to revise when occluded by tumor overgrowth. To compensate for such shortcomings, we developed a novel biliary metal stent with coil-spring structure. The aim of this study was to investigate the feasibility and efficacy of the novel stent in a swine model.

METHODS The biliary stricture model was prepared in six mini-pigs using endobiliary radiofrequency ablation. Conventional PS (n=2) and novel stents (n=4) were deployed endoscopically. Technical success was defined as successful stent placement and clinical success was defined as >50% reduction of serum bilirubin level. Adverse events, stent migration, and endoscopic removability for one month after stenting were also assessed.

RESULTS The biliary stricture was successfully created in all animals. The technical success rate was 100 %, and the clinical success rate was 50% in the PS group and 75% in the novel stent group. In the novel stent group, the median pre- and post-treatment serum bilirubin levels were 3.94 and 0.3 mg/dL. Stent migration occurred in two pigs and two stents were removed by endoscopy. There was no stent-related mortality.

CONCLUSIONS The newly designed biliary metal stent was feasible and effective in a swine biliary stricture model. Further studies are needed to verify the usefulness of the novel stent in the management of biliary strictures.

Copyright © 2023 Cho, Lee, Choi, Park, Lee, Kim, Jeong, Lee, Jeong, Ki, Paik, Ryu and Kim.

Keywords

animal experimentbile duct obstructionbiliary tract neoplasmsendoscopic retrograde cholangiopancreatographystent

Figures

Figure 1
Figure 1 Novel biliary metal stent with coil-spring structure. (A) Structure of novel stent. (B) Coil-spring structure was adopted to increase flexibility and prevent migration. (C) A silicone skirt was installed at the distal end to prevent duodenal mucosa damage and a lasso was attached for endoscopic removal. (D) Silicone coating was applied to both sides of the nitinol.
Figure 2
Figure 2 Fluoroscopic images of the endoscopic deployment of a novel biliary metal stent in a phantom model.
Figure 3
Figure 3 Endoscopic retrograde cholangiograms of animals. (A) Cholangiogram obtained during endobiliary RFA. RFA catheter positioned in the CBD (B) Cholangiogram at four weeks after endobiliary RFA showing stricture at site of procedure (C) Novel biliary metal stent was deployed successfully. Distal tip located in the duodenal bulb. (D) Stent was deployed across the stricture site.
Figure 4
Figure 4 The cross section of plastic stent (A) and novel biliary metal stent (B) were checked after endoscopic stent removal. Both stents had been placed for 4 weeks. Novel biliary metal stent showed a large inner diameter through less formation of biofilm and biliary sludge.
Figure 5
Figure 5 Histologic examination of bile duct. (A) Gross inspection of extracted bile duct was conducted. For microscopic examination, fixed sample was sectioned at 3 levels (black dots). (B, C) Inflammatory cell infiltration sites were seen in the bile duct mucosa of subject #2 and #3 (black arrow, H&E X100) (D) Inflammation scores of each subject were calculated and compared.

그림 출처: PubMed Central 공개 논문 (PMC9982730)

Conflict of interest statement

B-YK is employed by CGBio, a stent manufacturer. The remaining authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.