SCI 발표논문

SCI 발표논문

← SCI 발표논문 목록

Sci Rep. 2020;10(1):19015.

Validation of a wearable cuff-less wristwatch-type blood pressure monitoring device.

, , , , ,

제1저자 문준호, 강명균 · 교신저자 임수

소속 (6)
  1. Department of Internal Medicine, Seoul National University College of Medicine, Seoul, South Korea.
  2. Department of Internal Medicine, Seoul National University Hospital, Seoul, South Korea.
  3. InBody Co., Ltd, Seoul, South Korea.
  4. Department of Internal Medicine, Seoul National University Bundang Hospital, Seongnam, South Korea.
  5. Department of Internal Medicine, Seoul National University College of Medicine, Seoul, South Korea. limsoo@snu.ac.kr.
  6. Department of Internal Medicine, Seoul National University Bundang Hospital, Seongnam, South Korea. limsoo@snu.ac.kr.

Abstract

Ambulatory blood pressure (BP) monitoring is recommended to improve the management of hypertension. Here, we investigated the accuracy of BP estimated using a wearable cuff-less device, InBodyWATCH, compared with BP measured using a manual sphygmomanometer. Thirty-five adults were enrolled (age 57.1 ± 17.9 years). The BP was estimated using InBodyWATCH with an individualized estimation based on a neural network model. Three paired sets of BPs from the two devices were compared using correlation analysis and Bland-Altman plots (n = 105 paired BP readings). The correlations for both systolic and diastolic BP (SBP and DBP) between the two devices were high (r = 0.964 and 0.939, both P < 0.001). The mean difference was 2.2 ± 6.1 mmHg for SBP and -0.2 ± 4.2 mmHg for DBP; these were not significant (P = 0.472 for SBP and P = 0.880 for DBP). The proportions of estimated SBP/DBP obtained from the InBodyWATCH within ± 5 mmHg of manual SBP/DBP were 71.4%/83.8%; within ± 10 mmHg they were 86.7%/98.1%; and within ± 15 mmHg they were 97.1%/99.0%. The estimated BP from this wearable cuff-less device correlated highly with the manual BP and showed good accuracy, suggesting its potential to be used in ambulatory BP monitoring.

Figures

Figure 1
Figure 1 Wearable device (InBodyWATCH) and its application. (a) An adhesive electrode was attached to the right wrist to acquire ECG signals. (b) ECG signals was collected by putting a right-hand finger on the wristwatch. (c) An example of data display generated from the wearable device in the iOS system.
Figure 2
Figure 2 Correlation of the accuracy of systolic blood pressure (SBP) measurements between the InBodyWATCH (SBP_Watch) and a manual sphygmomanometer (SBP_Auscultatory).
Figure 3
Figure 3 Correlation of the accuracy of diastolic blood pressure (DBP) measurements between the InBodyWATCH (DBP_Watch) and a manual sphygmomanometer (DBP_Auscultatory).

그림 출처: PubMed Central 공개 논문 (PMC7642418)

Conflict of interest statement

The authors declare no competing interests.

그 밖의 정보
출판 형태
Journal Article, Research Support, Non-U.S. Gov't, Validation Study
저널
Scientific reports · eISSN 2045-2322
전자 출판
2020-11-04
MeSH
Adult · Aged · Algorithms · Blood Pressure Monitoring, Ambulatory / instrumentation · Calibration · Electrocardiography · Female · Humans · Male · Middle Aged · Pulse Wave Analysis · Sphygmomanometers · Wearable Electronic Devices* · Wrist*
참고문헌
33편